| OPTIZEN 3220UV |
| Double beam UV/Vis spectrophotometer |
|
 |
|
OPTIZEN 3220UV is very much ideal for analyzing components that require multiple wavelengths, time course operations, and getting stable data sheets. |
 |
 |
 |
 |
|
| |
| Features |
| |
- More powerful in information technologies in your lab.
- Better in holding memories.
- USB ports for user-friendly programming updates and other accessory connections.
- Universal sample cell holder holds 8 cells.
- Rotating cell holder (8 cells and one blank)
- Superb optics for maximized accuracy.
- Exceptional precision and reliability.
- User-friendly software with scanning and time course capacities.
- Versatile data processing.
|
|
| |
| OPTIZEN View 4.2 (PC interface software) |
| |

|
ATC mode
Measuring absorption, transmittance, and concentration. |
 |
STC mode
Draw standard curve. |
 |
SUR mode
Making graph at the specific wavelength ranges. |
 |
KIN mode
Measurement with variable with time differences |
|
| |
| Specifications |
| |
Optical System |
Double beam optics |
| |
Spectral Band
Pass Width(nm) |
< 1.0 |
| |
accuracy(nm) |
<±0.3 (at D2 Peak 656.1nm, 486.0nm) |
| |
reproducibility(nm) |
<±1.0 |
| |
slew rate(nm/min) |
about 5000 |
| |
scanning speed(nm/min) |
maximum 3000 |
| |
accuracy(Abs) |
±0.004 (at 1.0), ±0.002 (at 0.5) |
| |
reproducibility(Abs) |
±0.002 (at 1.0), ±0.001 (at 0.5) |
| |
stray light(%T) |
0.05 (220nm, 340nm) |
| |
Baseline flatness(ABS/h) |
<±0.001 (at 550nm) |
| |
Baseline Flatness(ABS) |
<± 0.001 (200 ~ 1100nm) |
| |
Drift(A/hr) |
0.002 (at 340nm) |
| |
Monochomater |
Modified Czerny-Turner type with 1200 lines/mm blazed Grating |
| |
Light Source |
Tungsten-Halogen Lamp & Deuterium Lamp |
| |
Lamp Change Wavelength |
340~410 (Default 370), include Auto Position System |
| |
Standard Cell Holder |
Automatic Rotary type 8-position Multi-Cell Holder |
| |
Interface Ports |
USB, Serial ports, Parrallel ports |
| |
Power Requirement |
Free Voltage |
| |
Overall
Dimensions(mm) |
600(W)x400(D)x250(H)
|
| |
Weight |
16kg(Only Main System) |
|
| |
Accessories  |
| |
Temperature cell holder
Micro volume cell holder
Film cell holder–Wide
microvolume cell holder
|
|
| |
| Pre-programmed products (3220 series) |
| |
Nanobiometer OPTIZEN 3220 bio
Water quality analyzer OPTIZEN 3220 QX  |
| ※Please contact our center overseas@mecasys.co.kr |
| |
| |